Canon TX-F Spécifications Page 60

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 78
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 59
8.2 Before Calling a Service Person
– 50 –
* xxx indicates a three-digit number.
Error in the system
LOST BACKUP DATA
Problem has occurred in the
instrument.
Although you can continue the
measurement for time being, the
data cannot be backed up.
Immediately ask Canon or dis-
tributor for replacement.
LOST DATE DATA
LOW BATTERY
Voltage of the battery for data
backup is low.
Although you can continue the
measurement for time being, the
data that are backed up will be
lost in time. Immediately ask
Canon or distributor for replace-
ment.
SYSTEM ERROR
(xxx*)
Problem has occurred in the
instrument.
Turn OFF the power switch and
turn it ON again after about 10
seconds.
ILLEGAL CODE ERROR
Problem has occurred in the
instrument.
Turn OFF the power switch and
turn it ON again after about 10
seconds.
ILLEGAL SLOT ERROR
CPU ADDRESS ERROR
DMAC ADDRESS ERROR
Message / Display Cause Remedy